Z-2317-2024: GE Vivid Ultrasound products, Model Numbers: a) Vivid S70, H45041SW

Recall record Z-2317-2024 for GE Vingmed Ultrasound As. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2317-2024
Company in FDA record
GE Vingmed Ultrasound As
FDA classification date
Jul 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.

Source facts

Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 05/02/2024 via letter delivered by traceable means. The notice explained the problem with the device, potential safety issue, and requested the following: "Actions to be taken by Customer/User: You can continue to use your device in accordance with the User Manuals, specifically: 1. After selecting a patient in the DICOM Modality Worklist, make sure to review and confirm that the patient demographics shown on the ultrasound scanner screen during the exam are from the intended patient. 2. In the scenarios where the incorrect and/or actual patient already exists in the patient archive, a Patient Match dialog can be shown to inform you about patient information inconsistency. Make sure to review the patient demographics properly. 3. To reduce the probability of occurrence of the issue, reduce the size of the DICOM Worklist search by: a) Setting Max. Results to 75 in the Worklist dataflow Config dialog (available by selecting Config | Connectivity | Dataflow | Worklist* | Inputs | DicomWorklist | Properties) and/or b) Configuring the DICOM Worklist server to automatically delete Scheduled Procedure Steps that have been performed on the ultrasound scanner. If this option is not available on your DICOM Worklist server, make sure to manually clear all Scheduled Procedure Steps that have been performed from the DICOM Worklist on a regular basis (e.g., every day). *The actual instructions on how to perform this step depends on which DICOM Worklist server is used in your department. Consult your IT department to find out how to perform the step. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions." "GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction."
Initiated
2024-05-02
Root cause
Device Design
Distribution
US: IL, MA, NM, NY, OK,
Recall number
Z-2317-2024
Recall status
Open, Classified
CDRH record ID
208236
FDA classified
2024-07-05
Recall event number
94772

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗