Z-2318-2008: Universal SlingBar 350 with Quick-release Hook Product Number: 3156084 Used in c

Recall record Z-2318-2008 for Liko AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2318-2008
Company in FDA record
Liko AB
FDA classification date
Sep 20, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A component securing the Universal SlingBar to the patient lift, may experience a nut unthreading resulting in the sling bar detaching from the lift causing the patient to fall.

Source facts

Action
Liko North America ,the US Distributor, notified accounts by letter dated June 30, 2008 and provided a Field Safety Notice with Safety Inspection Procedures to performed on the lift prior to use. Replacement slingbars are being provided. If you have questions, contact the Liko product distributor in your country. Contact information can be found at www.liko.com/int/distributors.asp.
Initiated
2008-07-01
Root cause
Component design/selection
Terminated
2015-07-17
Distribution
Nationwide
Recall number
Z-2318-2008
Recall status
Terminated
CDRH record ID
72052
FDA classified
2008-09-20
Recall event number
48747

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗