Skip to content

Public FDA record

Z-2318-2024: 9006767 Astra EV 4.8mm Milled Titanium Abutment Milled abutments are too be used in conjunction with an

Recall record Z-2318-2024 for Preat Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2318-2024
Company in FDA record
Preat Corp
FDA classification date
Jul 8, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to incorrect milled titanium abutment being packaged and labeled.

Source facts

Action
On 05/21/2024, the firm sent via direct certified mail an "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter to customers informing them that product labeled as part number 9006767-F Astra EV 4.8mm Milled Titanium Abutment may contain instead part number 9006667 Astra EV 4.2mm Milled Titanium Abutment. The potential risk identified was, the screw seat is higher than specified in the engineering drawings and required for full thread engagement with the implant. This condition reduces the screw thread engagement with the implant. Customer are instructed to: Product in the scope of this recall shall be discarded or returned to PREAT CORPORATION. Instructions for the return will be provided. All personnel within your company should be made aware of the issue. Any customers to which this product has been further distributed should be notified. For questions or assistance contact Jose Espino at 224-622-7191 or email jespino@younginnovations.com
Initiated
2024-04-10
Root cause
Process control
Distribution
U.S.: CA, MO, and NV. O.U.S.: None
Recall number
Z-2318-2024
Recall status
Open, Classified
CDRH record ID
208009
FDA classified
2024-07-08
Recall event number
94693

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗