Skip to content

Public FDA record

Z-2319-2013: DeVilbiss Vacu-Aide QSU Suction Unit (Disposable Suction Container/Filter), 800 ML DISP Container, Unasse

Recall record Z-2319-2013 for DeVilbiss Healthcare LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2319-2013
Company in FDA record
DeVilbiss Healthcare LLC
FDA classification date
Oct 1, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

End users of the DeVilbiss Disposable Suction Container/Filter had exposed the bacterial filter cartridge to fluid, thereby occluding the filter and rendering it unusable.

Source facts

Action
DeVilbiss Healthcare sent End users a field corrective action letter dated March 7, 2013, to all affected customers to reinforce the proper cleaning method. This letter is to inform you that we have come across cases where end users had exposed the bacterial filter cartridge in the disposable suction container to fluid, thereby occluding the filter and rendering it unusable Specifically what do you have to do when you receive a correction/recall notice? " The US FDA states in its correction/recall regulations that: "Consignees that receive a recall communication should immediately carry out the instructions set forth by the recalling firm and, where necessary, extend the recall to its consignees." "Enclosed you will find two Urgent Medical Device Correction letters addressed to DeVilbiss Dealer/Provider" and "DeVilbiss User" that we ask you read carefully. "It is important that you follow the actions outlined in these letters to ensure we can effectively disseminate this information to end-users. "To reach your direct accounts we ask that you send a copy of the "DeVilbiss User" letter to each of these end-users. We sincerely appreciate all of your efforts to ensure the safety of your customers that depend on DeVilbiss products. We thank you in advance for your assistance and apologize for any inconvenience that you may experience as a result of this situation. Should you have any questions or concerns, please contact DeVilbiss Healthcare Customer Service at 1-800-338-1988.
Initiated
2013-03-07
Root cause
Use error
Terminated
2014-11-06
Distribution
Worldwide Distribution - USA and the countries of Mexico and Canada.
Recall number
Z-2319-2013
Recall status
Terminated
CDRH record ID
122122
FDA classified
2013-10-01
Recall event number
66347

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.