Open original FDA record ↗Recall record
- FDA identifier
- Z-2319-2019
- Company in FDA record
- Qiagen Sciences LLC
- FDA record creation date
- Aug 21, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
QIAsymphony software version 5.0.3 - Software issue incorreclty assigns the 2D bar code eluate ID to the sample within the results file. If 2D bar code eluate IDs are not checked against the sample ID prior to downstream application the incorrect tube may be selected and it could have the potential to lead to delayed results or incorrect patient reporting
Source facts
- Action
- Qiagen notified Customer on July 2, 2019 via email and provided reason for recall, heath risk and action to take: Discontinue use of the workflow in the following specific condition: QIAsymphony software version 5.0.3 when using the 2D bar code integration feature with continuous loading. Please note that the 2D bar code labware, continuous loading, and Software 5.0.3 can still be used independently of the integration feature. Complete and sign the included Acknowledgement of Receipt form and email it to QIAGEN Technical Services at techservice-eu@qiagen.com or fax it to +49 (0)2103-29-22400. Questions, please contact QIAGEN Technical Services Department. Telephone: 800 362 7737. Email: TechService-NA@qiagen.com
- Initiated
- 2019-07-02
- Root cause
- Software design
- Distribution
- Worldwide Distribution - US Nationwide Foreign: Algeria Australia Austria Azerbaijan Belgium Brazil Canada China Czech Republic Denmark Finland France Germany Greece India Israel Italy Japan Jordan Korea, Republic of Kuwait Latvia Mexico Netherlands Norway Pakistan Poland Republic of Korea Russian Federation Singapore South Africa Spain Sweden Switzerland Thailand Turkey United Kingdom
- Recall number
- Z-2319-2019
- Recall status
- Open, Classified
- CDRH record ID
- 175105
- FDA record created
- 2019-08-21
- Recall event number
- 83409