Z-2319-2024: Impella CP with SmartAssist, Product No.: 1000080 (REF 0048-0003). left heart support blood pump

Recall record Z-2319-2024 for Abiomed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2319-2024
Company in FDA record
Abiomed, Inc.
FDA classification date
Jul 17, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Nine (9) Impella CP pumps failed inspection and were inadvertently released.

Source facts

Action
On May 31, 2024 consignees were contacted by phone with a script followed-up by sending URGENT VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) letters to consignees. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER: 1.Examine your inventory immediately to determine if you have product subject to this recall(removal). QUARANTINE AND DO NOT USE THE SUBJECT PRODUCTS. 2.Your Abiomed Representatives have been provided instructions to support facilitation of locating, separating, and returning affected product for replacement. Work with your Abiomed Representative toreview, complete all fields, and sign the Business Response Form attached on page 4. IMPORTANT:The Business Response Form must be completed even if you have used any units of the subjectproduct on hand. 3.Forward this notice to any personnel in your facility who need to be informed. 4.If any of the product subject to this recall (removal) has been forwarded to another facility, contact thatfacility and provide a copy of this notice to the relevant personnel. 5.Post a copy of this notice in a visible area for awareness of this field safety notice. 6.As with any medical device, adverse reactions or quality problems experienced with the use of this productshould be reported to the FDA s MedWatch Adverse Event Reporting Program using the link below: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting- program/reporting-serious-problems-fda Please contact Abiomed Customer Service, 1-800-422-8666, option 2 (email recall@abiomed.com or contact your local clinical field staff if you have questions or concerns regarding this notice.
Initiated
2024-05-31
Root cause
Release of Material/Component prior to receiving test results
Distribution
Domestic only: FL, MA, OH TX.
Recall number
Z-2319-2024
Recall status
Open, Classified
CDRH record ID
208475
FDA classified
2024-07-17
Recall event number
94833

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗