Skip to content

Public FDA record

Z-2321-2026: Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;

Recall record Z-2321-2026 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2321-2026
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Jun 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.

Source facts

Action
On 5/12/2026 an IMPORTANT MEDICAL DEVICE INFORMATION letters were sent to customers to inform them that manipulation of the luer-lock location during treatment preparation may result in fluid spillage onto the user and floor. This occurs prior to treatment and renders the product unusable. Direct contact with the fluid does not present any risk to the user as frangible manipulation only occurs after the peal seam has been broken and the two fluid compartments have been mixed. The potential risk associated with the affected Bicarby Dialysate product is a slip and fall hazard due to the presence of fluid on the floor.
Initiated
2026-05-12
Root cause
Process design
Distribution
US Nationwide distribution in the states of NY, TN and TX.
Recall number
Z-2321-2026
Recall status
Open, Classified
CDRH record ID
220066
FDA classified
2026-06-05
Recall event number
98877

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗