Open original FDA record ↗Recall record
- FDA identifier
- Z-2322-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Sep 6, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
During regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system. During normal clinical operation the switch on the motor drive handle of the unit may become loose and slide horizontally to the right and become trapped under the cover. This may cause the system to continuously move. In response, Siemens has issued Update Instruction XP038/12/S to r
Source facts
- Action
- Siemens initiated their correction of the MOBILETT Mira on August 21, 2012 by sending an "UPDATE INSTRUCTIONS XP038/12/S to all affected customers. The Instruction identifies the product, problem, and work steps for customers to follow. A XP038/12/S Completion Protocol letter was included for customers to complete and file with the instruction once the update is completed.
- Initiated
- 2012-08-21
- Root cause
- Device Design
- Terminated
- 2014-04-03
- Distribution
- Nationwide Distribution- including the states of GA, ID, IL, IA, KS, NC, OH, SD, TX, and WI.
- Recall number
- Z-2322-2012
- Recall status
- Terminated
- CDRH record ID
- 112355
- FDA classified
- 2012-09-06
- Recall event number
- 63042