Open original FDA record ↗Recall record
- FDA identifier
- Z-2322-2016
- Company in FDA record
- Zimmer Gmbh
- FDA classification date
- Aug 2, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Zimmer Biomet initiated a voluntary recall of the Twist Drill & V-Tek Twist Drill due to to a potential tolerance stack up identified between the twist drill and the mating Kirschner Wire (K-Wire) component.
Source facts
- Action
- Zimmer GmbH initiated a voluntary recall of the Normed - TWIST DRILL Cannulated & Normed- V-TEK TWIST DRILL Cannulated, with written notices on June 27, 2016 via certified mail due to a potential tolerance stack up identified between the twist drill and the mating Kirschner Wire (K-Wire) component. Customers are asked to do the following: Review the notification and ensure affected personnel are aware of the contents. 2. Locate all affected product identified above and quarantine them immediately. 3. Return any affected product within your possession. Clearly mark the outside of all return packages, Recall, and include a copy of the Inventory Return Certification form (Attachment 1) with your return shipment(s). 4. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-800-348-2759 between 8:00 am and 8:00 pm EST. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. " Online:www.fda.gov/medwatch/report.htm " Regular Mail: use postage-paid, pre-addressed Form FDA 3500 available at: www.fda.gov/MedWatch/getforms.htm. Mail to address on the pre-addressed form. " Fax: 1-800-FDA-0178 Affected units returned as part of this recall activity will be scrapped
- Initiated
- 2016-06-27
- Root cause
- Under Investigation by firm
- Terminated
- 2017-05-30
- Distribution
- Domestic: CO, IL Foreign: Argentina, Costa Rica, India, Malta, Japan, Philippines, Vietnam, Portugal, Dominican Republic, Argentina, Australia, Russia, Saudi Arabia, Ireland, Ecuador, China, Singapore, Indonesia, Canada, Mexico, South Korea , Spain, Netherlands, Sweden, Malaysia, New Zealand, Switzerland, El Salvador, England, Norway, Finland, Venezuela, Taiwan, Italy, Colombia, Greece, Thailand, France, Israel, Jamaica, Japan, Jordan, Germany, Korea, Lebanon, Belgium, Hong Kong, Philippines, Iran, Brazil, Turkey, United Arab Emirates (UAE), Uruguay
- Recall number
- Z-2322-2016
- Recall status
- Terminated
- CDRH record ID
- 147880
- FDA classified
- 2016-08-02
- Recall event number
- 74548