Z-2322-2024: Volara System. Blue Ventilator Adapter Module. Intermittent positive pressure breathing device component

Recall record Z-2322-2024 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2322-2024
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Jul 17, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.

Source facts

Action
Baxter Healthcare notified customers on about 05/30/2024 via letter sent first-class mail. The letter instructed facilities to immediately locate, isolate, and cease all use of the affected lot numbers, return affected units on hand, and complete and return the acknowledgement form. If the product was further distributed, they were instructed to notify customers if product was further distributed. The letter instructed patients to immediately locate, isolate, and cease all use of the affected lot numbers, return affected units on hand, however if they do not have other options in stock, in order to not delay therapy, you may continue to use with caution. Visually inspect the nebulizer port prior to therapy and assess for ventilator gas flow leaks for the duration of therapy. They were also instructed to complete and return the acknowledgement form.
Initiated
2024-05-30
Root cause
Device Design
Distribution
Nationwide distribution. International distribution to Canada, France, and Germany.
Recall number
Z-2322-2024
Recall status
Open, Classified
CDRH record ID
208119
FDA classified
2024-07-17
Recall event number
94713

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗