Z-2322-2025: Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860

Recall record Z-2322-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2322-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Aug 13, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or inappropriate therapy adjustments in patients being monitored for residual or recurrent thyroid cancer.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification dated 7/2/25 was sent to customers. ACTION: " Discontinue using Access Thyroglobulin reagent lot 439163 and discard all remaining reagent packs from this lot. " At the discretion of the medical director, a retrospective review of patient results generated with Access Thyroglobulin reagent lot 439163 should be performed, which includes reviewing patient history, to assess the need for retesting. " Please contact your local Beckman Coulter representative for replacement product requests and to receive updates on availability and delivery. RESOLUTION: Beckman Coulter is investigating the root cause of this issue to prevent similar occurrences. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via email. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center: " From our website: http://www.beckmancoulter.com " Contact your local Beckman Coulter Representative for replacement.
Initiated
2025-07-10
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide and the countries of Brazil, Canada, Georgia, Italy, Malaysia, Mexico, Philippines, Spain.
Recall number
Z-2322-2025
Recall status
Open, Classified
CDRH record ID
214974
FDA classified
2025-08-13
Recall event number
97255

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗