Skip to content

Public FDA record

Z-2323-2021: Cytocell LPH019-A E2A (TCF3) Breakapart Probe- In Vitro Diagnostic Analyte specific reagents (ASR's)

Recall record Z-2323-2021 for Cytocell Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2323-2021
Company in FDA record
Cytocell Ltd.
FDA record creation date
Aug 18, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

May show unexpected locus specific signals in addition to those at 19p13.3. Users may observe faint additional locus specific signals at 9q34

Source facts

Action
Cytocell issued Urgent Medical Device Recall Users issued via email on 5/27/21. Letter states reason for recall, heath risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that you destroy the remaining inventory. We also suggest that laboratories undertake a review any results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of any additional locus specific signal at 9q34. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Initiated
2021-05-27
Root cause
Nonconforming Material/Component
Terminated
2022-05-09
Distribution
US Nationwide distribution in the states of CA, IL, KY, NY, OH, TX, UT, VA.
Recall number
Z-2323-2021
Recall status
Terminated
CDRH record ID
188305
FDA record created
2021-08-18
Recall event number
88294

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗