During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.
Source facts
Action
On 06/03/2024, Baxter Healthcare's Customer Service team contacted the consignee, via telephone call, to discuss the recall action and request the affected unit be returned to Baxter.
Initiated
2024-06-03
Root cause
Employee error
Distribution
Distribution to Illinois.
Recall number
Z-2323-2024
Recall status
Open, Classified
CDRH record ID
208382
FDA classified
2024-07-10
Recall event number
94805
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.