Z-2324-2024: ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed

Recall record Z-2324-2024 for Zoll Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2324-2024
Company in FDA record
Zoll Medical Corporation
FDA classification date
Jul 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest

Source facts

Action
Zoll Medical issued Urgent Corrective Distributor notification on 31-May-24 and consignees notification commenced 07-Jun-24 via email. Letter states reason for recall, health risk and action to take: Customers who have affected devices should immediately take the following steps: (1) Alert G5 Semi-Automatic AED users of this problem. (2) Locate the affected devices. (3) Remove the protective film in accordance with Appendix A. (4) Respond to ZOLL via the customer notification form acknowledging that the protective film has been removed. Technical support numbers 1 (800) 348-9011 or 1 (978) 421-9460 and are available to assist users with any aspect of this notice.
Initiated
2024-05-31
Root cause
Use error
Distribution
Nationwide including PR. Foreign: AU BM CA CH DE DK ES FR GB HK HU ID IE IL IT KR MA MT NL NO PR SA SE SI SN TR
Recall number
Z-2324-2024
Recall status
Open, Classified
CDRH record ID
208336
FDA classified
2024-07-10
Recall event number
94792

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗