Open original FDA record ↗Recall record
- FDA identifier
- Z-2324-2025
- Company in FDA record
- Diversatek Healthcare
- FDA classification date
- Aug 14, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Anorectal manometry procedure kit outer kit label is mislabeled with an incorrect expiration date; the manufacturing date was used in place of the expiration date. All inner labels on kit components display the correct expiration date.
Source facts
- Action
- On 6/4/2025, field safety notices were emailed to GI Lab / Endoscopy Unit Managers who were asked to do the following: Check your inventory for devices with the part and lot numbers listed below. 2. Discard any affected product. 3. Complete and return the attached MEDICAL DEVICE RECALL RETURN RESPONSE form via email to clahr@diversatek.com If you have questions about this issue, please contact your Sales Representative or the firm at 1-800-558-6408
- Initiated
- 2025-06-04
- Root cause
- Error in labeling
- Distribution
- Worldwide - US Nationwide distribution in the states of NM, MA, CO, OH and the countries of Mongolia, South Korea.
- Recall number
- Z-2324-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215061
- FDA classified
- 2025-08-14
- Recall event number
- 97282