In the system (bk3500) software generic volume measurement (HxWxLxl) function the user may be
unaware of the calculation factor (1) and assume the generic volume calculation factor is similar to that of
the other (prostate, adenoma, testis, kidney, and bladder) volume calculation factors provided in the
software.
Source facts
Action
BK Medical sent an Important Corrective Action Notification letter dated June 27, 2016, to all affected customers. The letter explains the issue and asks customers to discontinue the use of the generic volume measurement and contact BK medical ApS Service for assistance. Customers with questions were instructed to call 800-645-4442.
Initiated
2016-06-28
Root cause
Software design
Terminated
2016-08-29
Distribution
Worldwide Distribution - US to TX, FL, and MA. Internationally to Australia
Recall number
Z-2325-2016
Recall status
Terminated
CDRH record ID
148228
FDA record created
2016-08-03
Recall event number
74691