Skip to content

Public FDA record

Z-2325-2024: Kosmos on iOS impacting iOS, Model Numbers: P007639, P007944, software is part of a general-purpose diagn

Recall record Z-2325-2024 for Echonous Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2325-2024
Company in FDA record
Echonous Inc
FDA classification date
Jul 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ultrasound system has a bug in affected iOS software, which after DICOM export, can cause inaccuracy in postprocessed measurements done in 3rd party software, such as PACS or other post-processing software, which may lead to inaccurate diagnoses or treatment.

Source facts

Action
On 5/9/24, notices were sent to customers informing them to refrain from performing measurements on Kosmos iOS images with third-party PACS or post-processing software, after DICOM export. Also, they were informed, that a software patch would be ready to upload soon. On 5/22/24, notices were sent to customers informing them that software patch 2.1.4.53 was available and that users should download this software performing measurements on Kosmos iOS images with third-party PACS or post-processing software. Follow the steps below and download this new version as soon as possible: 1. Open your currently installed Kosmos Ultrasound app and from the home screen go to "Settings" > "DICOM". Write down your MWL and PACS profiles. You will need to reference these profiles after you download the new app. Please note: your MWL and PACS profiles will not be retained when you download the new app. If you don't use DICOM or MWL then you can skip this step. 2. Export any images or clips you wish to keep via USB or DICOM. Please note: existing images or clips will not be retained when you download the new app. 3. Delete the Kosmos Ultrasound app that is currently on your iPad(s) 4. Download Kosmos Ultrasound 2.1.4.53 from the Apple App store 5. Connect your iPad(s) to Wi-Fi and open the Kosmos Ultrasound 2.1.4.53 app to download any purchased licenses and when prompted enable drivers 6. From the home screen go to "Settings" > "DICOM" and reconfigure your MWL and PACS profiles with the profile information you wrote down in step 1. If you don't use DICOM or MWL then you can skip this step also. Fill out and return the acknowledgement form. Questions or concerns can be directed to Customer Service at (844) 854-0800 service@echonous.com
Initiated
2024-05-09
Root cause
Software Design Change
Distribution
US: ID, TX, WI, KS, IL, GA, MI, CA, OH, FL, AK, UT, MD, NY, NV, CO, ND, WA, DC, LA, KY, MN, HI, NM, TN, NJ, ID, MI, VT, AZ, MO, AL, MA, NC OUS: Denmark, Switzerland, Austria, Italy, France, United Kingdom, Germany, Spain, Romania, Finland, Romania, Australia, India, Hong Kong, Thailand
Recall number
Z-2325-2024
Recall status
Open, Classified
CDRH record ID
208061
FDA classified
2024-07-10
Recall event number
94709

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗