Open original FDA record ↗Recall record
- FDA identifier
- Z-2326-2012
- Company in FDA record
- Breg Inc, An Orthofix Company
- FDA classification date
- Sep 6, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
This recall has been initiated due to confirmation that the Polar Dressing, Knee/MU/Shldr Intelli-Flo-Breg does not meet the required thickness as specified. The dressing in the sterile pouch measures a thickness of 0.010" instead of 0.020". This change in thickness makes the dressing thinner, less durable and subject to tearing. The product is mislabeled.
Source facts
- Action
- The firm, Breg Orthofix, called their customers on August 11, 2010 sent a letter dated August 12, 2010 and emails to all customers. The call and letter described the product, problem and actions to be taken. The customers were instructed to take immediate action to identify and quarantine the product at their facility. Breg established a return materials authorization and is contacting their customers to provide a Return Authorization to return the dressings. Breg is also providing replacement dressings. If you have any questions regarding communication or the Intelli-Flo Sterile Dressing, contact Breg Customer Relations at (800) 321-0607.
- Initiated
- 2010-08-11
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-09-06
- Distribution
- Worldwide distribution: USA (nationwide) and country of: Nova Scotia.
- Recall number
- Z-2326-2012
- Recall status
- Terminated
- CDRH record ID
- 98803
- FDA classified
- 2012-09-06
- Recall event number
- 58247