Skip to content

Public FDA record

Z-2326-2012: Polar Dressing, Knee/MU/Shldr Intelli-Flo-Breg, PN 10630 Used with cold therapy pads to provide insula

Recall record Z-2326-2012 for Breg Inc, An Orthofix Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2326-2012
Company in FDA record
Breg Inc, An Orthofix Company
FDA classification date
Sep 6, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

This recall has been initiated due to confirmation that the Polar Dressing, Knee/MU/Shldr Intelli-Flo-Breg does not meet the required thickness as specified. The dressing in the sterile pouch measures a thickness of 0.010" instead of 0.020". This change in thickness makes the dressing thinner, less durable and subject to tearing. The product is mislabeled.

Source facts

Action
The firm, Breg Orthofix, called their customers on August 11, 2010 sent a letter dated August 12, 2010 and emails to all customers. The call and letter described the product, problem and actions to be taken. The customers were instructed to take immediate action to identify and quarantine the product at their facility. Breg established a return materials authorization and is contacting their customers to provide a Return Authorization to return the dressings. Breg is also providing replacement dressings. If you have any questions regarding communication or the Intelli-Flo Sterile Dressing, contact Breg Customer Relations at (800) 321-0607.
Initiated
2010-08-11
Root cause
Nonconforming Material/Component
Terminated
2012-09-06
Distribution
Worldwide distribution: USA (nationwide) and country of: Nova Scotia.
Recall number
Z-2326-2012
Recall status
Terminated
CDRH record ID
98803
FDA classified
2012-09-06
Recall event number
58247

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗