Skip to content

Public FDA record

Z-2326-2024: Decompression Table Model E9011. This device applies cervical and/or lumbar decompression with practition

Recall record Z-2326-2024 for Pivotal Health Solutions, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2326-2024
Company in FDA record
Pivotal Health Solutions, Inc.
FDA classification date
Jul 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This product is being recalled due to potentially insufficient rework on a part (E7130-B) on the lumbar section of the frame on 6 distributed tables. There were 4 additional tables with the faulty E7130-B part that were manufactured, but had not been distributed yet.

Source facts

Action
Consignees were initially notified of the recall via phone call starting on 4/29/2024. A followup email or certified mail were sent to the consignees on 6/11/2024. Customers were instructed that Fed Ex was contacted for pickup of the table so the table can be packaged and returned to Pivotal Health Solutions at 3003 9th Ave. SW, Watertown, SD 57201. A recall expansion led to a second updated recall notification being sent to consignees beginning on June 26, 2024 via email. This letter instructs the consignees to stop using the tables listed in the letter. The firm will be contacting the consignees to schedule an engineer to come out to the consignee site to inspect the table(s). The table(s) will be replaced in found to contain an affected part. The consignee is also asked to return a response form.
Initiated
2024-04-29
Root cause
Nonconforming Material/Component
Distribution
Distributed to CA, FL, TX, ID, GA, and TN
Recall number
Z-2326-2024
Recall status
Open, Classified
CDRH record ID
208394
FDA classified
2024-07-10
Recall event number
94815

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗