Z-2327-2020: Infinix-8000F with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital r

Recall record Z-2327-2020 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2327-2020
Company in FDA record
Canon Medical System, USA, INC.
FDA record creation date
Jun 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.

Source facts

Action
On 03/03/20, Urgent Medical Device Correction notices were mailed to Radiology Managers/Directors. To correct this issue, the bolts holding the longitudinal movement of the table will need to be inspected to determine if the bolts are tightened to specification. Recalling firm service representatives will contact customers to set up appointments to inspect bolts. Radiology Managers/Directors were advised of the following: a) Should this problem occur before your appointment with the recalling firm's service representative, please stop using the system and contact the service representative. b) If abnormalities are found operating the table during a procedure, please stop use, and contact the service representative. c) Share this information with all users and reviewing radiologists as well as clinical engineering, and the biomedical group at your facility. d) Complete and return the Customer Reply Form. Additional questions should be directed to (800) 421-1968 or your local service representative (800) 521-1968
Initiated
2020-03-03
Root cause
Process control
Distribution
US Nationwide distribution including in the states of PA, AR, OH, NC, CA, IA, GA, IL, AL, WV, LA, MO, FL, MS, TX, AZ, MA, NJ, MD, NY, NV, OK, WI, PR.
Recall number
Z-2327-2020
Recall status
Open, Classified
CDRH record ID
181835
FDA record created
2020-06-09
Recall event number
85309

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗