Skip to content

Public FDA record

Z-2327-2024: Presource kits and trays, labeled as: 1) ANGIOGRAPHY PACK, Catalog Number SAN13BS605; 2) ANGIOGRAPHY P

Recall record Z-2327-2024 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2327-2024
Company in FDA record
Cardinal Health 200, LLC
FDA classification date
Jul 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cardinal Health is following the FDA's recommendations regarding the disposition of affected Jiangsu Shenli Medical Production Co. Ltd (Shenli) Syringes in the Merit Medical supplied finished kits. Merit Medical supplied finished kits are included as components in Presource kits. According to the FDA s communication, these Shenli syringes, located inside a Merit Medical kit component, cannot be distributed for use as they could contain a significant change or modification in design which could significantly alter the safety or effectiveness of the finished kit product, resulting in the risk of patient harm such as inaccurate dosing and a leaking device.

Source facts

Action
Cardinal Health issued an URGENT: PRODUCT RECALL notice to its consignees on 05/28/2024 via letter. The notice explained the issue, potential risks involved, and requested the following: "If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all syringes within the affected Merit Medical kit in the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: gmb-FieldCorrectiveAction@cardinalhealth.com For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits, please notify your customers of this recall." Expansion letter was sent via FedEx Express on 6/20/2024
Initiated
2024-05-29
Root cause
Device Design
Distribution
Nationwide
Recall number
Z-2327-2024
Recall status
Open, Classified
CDRH record ID
208356
FDA classified
2024-07-10
Recall event number
94763

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗