Skip to content

Public FDA record

Z-2329-2015: Puritan Bennett 980 Ventilator System, Universal (with neonatal functionality enabled) and Neonatal Model

Recall record Z-2329-2015 for Covidien LP (formerly Nellcor Puritan Bennett Inc.). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2329-2015
Company in FDA record
Covidien LP (formerly Nellcor Puritan Bennett Inc.)
FDA classification date
Aug 14, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Reports in which tidal volumes reaching patients were lower than set tidal volumes in neonatal Volume Control Plus (VC+) Mode with active humidification. This situation may potentially lead to respiratory compromise if not recognized.

Source facts

Action
Consignees will be notified via letter on July 17, 2015. No products will be returned, the reconfiguration of affected ventilators will occur at customer sites.
Initiated
2015-07-16
Root cause
Software design
Terminated
2016-06-02
Distribution
Worldwide Distribution-US (nationwide) and the countries of Canada, United Arab Emirates, Argentina, Australia, Bahrain, Brazil, Chile, Colombia, Costa Rica, Egypt, Ireland, Israel, Jordan, Japan, Kuwait, Lebanon, Morocco, Mexico, Panama, Peru, Qatar, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Turkey, and Uruguay.
Recall number
Z-2329-2015
Recall status
Terminated
CDRH record ID
138678
FDA classified
2015-08-14
Recall event number
71675

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗