Skip to content

Public FDA record

Z-2330-2008: Propper short gas-chex EO sterilization indicators, 250 strips per box, Reorder No. 269001, Propper Manuf

Recall record Z-2330-2008 for Propper Mfg Co Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2330-2008
Company in FDA record
Propper Mfg Co Inc
FDA classification date
Sep 20, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The firm discovered a production specification discrepancy that may cause the indicators to show an inaccurate result.

Source facts

Action
On 2/07/08, Propper Mfg. Co. sent URGENT: DEVICE RECALL letters (dated 2/07/08) to the direct consignees, informing them of the recall of the affected lots of Gas-Chex EO Indicator Strips. The letters also instructed them to: (1) check their inventory to determine if they have the recalled lots in stock; (2) discontinue use of the affected lots; (3) reprocess any unused packs that have been processed with the affected lots of Gas-Chex EO Indicator Strips with non-recalled lots or an alternative product from another supplier; (4) return the affected lots to Propper Manufacturing; (5) complete the enclosed form and return via fax to Propper Mfg. Propper Mfg. will send a postage-paid label and instruction for return of the product, and will ship replacement product as soon as possible. Propper also recommended that consignees review the risk associated with items that have been monitored with the subject chemical indicators although a sterilization cycle failure is unlikely due to equipment controls and other monitoring methods. For questions, please contact Propper at 1-800-832-4300 during regular business hours (9 am to 4:30 pm).
Initiated
2008-02-07
Root cause
Equipment maintenance
Terminated
2012-08-03
Distribution
Nationwide and Puerto Rico
Recall number
Z-2330-2008
Recall status
Terminated
CDRH record ID
72429
FDA classified
2008-09-20
Recall event number
48969

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗