Z-2330-2020: Mazor X Surgical System Positioner Type II, REF: ASM0214-02

Recall record Z-2330-2020 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2330-2020
Company in FDA record
Medtronic Navigation, Inc.
FDA classification date
Jun 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma, bleeding, or fracture(s)]

Source facts

Action
The firm originally sent a "Medical Device Correction " letter to affected consignees on December 5, 2019. On November 9, 2020, a "Update - Medical Device Correction" letter was sent out to all affected customers. The purpose of the letter was to update consignees that since the initiation of the recall (12/5/2019), Medtronic had developed a fix to permanently correct the issue associated with the Mazor X Surgical system. The notice further stated that upon availability of parts, Medtronic will schedule a service visit to correct the system at no charge. Medtronic expects to initiate these service visits within the next six months. Pending completion of service to correct the Positioner Type II system, continue to follow the modified mounting instructions. The firm requested that consignees do the following: 1. Please sign and date the Customer Confirmation Form enclosed with this letter to confirm that you have reviewed the information included in this notification with all users of the affected Mazor X" Surgical System, including physician users. 2. Return the form to Medtronic via email at RS.NavFCA@Medtronic.com or via fax at 651-367-7075 within 30 days of receipt. 3. If you have any questions regarding this notification, please contact Medtronic at 1-888-826-5603. Medtronic will initiate service visits to correct systems on-site at customer locations when the kits become available. This is expected to begin by February 2021. View spreadsheet for updated
Initiated
2019-12-05
Root cause
Device Design
Distribution
US: ALAR,AZ,CA,CO,CT,FL,GA,IA,ID,IL,IN,KY,MA,MI,MN,MO,NC,ND,NL,NY,OH,PA,SC,TN,TX,UT,VA,WA,WI,WV, OUS: Australia, China, India, Israel, Romania, Singapore, Thailand, United Kingdom, Switzerland, Spain, Germany, Belgium,
Recall number
Z-2330-2020
Recall status
Open, Classified
CDRH record ID
180727
FDA classified
2020-06-09
Recall event number
85387

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗