Z-2330-2021: AQUABEAM Handpiece, REF: HP2000, a sterile single-use component of the AQUABEAM Robotic System.

Recall record Z-2330-2021 for PROCEPT BIOROBOTICS CORPORATION. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2330-2021
Company in FDA record
PROCEPT BIOROBOTICS CORPORATION
FDA record creation date
Aug 20, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Scope tube tip may detach from the telescoping tube, which is attached to a single-use handpiece, a component of an aquablation robotic system. Both tip detachment and entire tip separation form the handpiece may lead to procedural delay for handpiece exchange. Tip detachment can be seen cystoscopically and tip separation from the handpiece may necessitate tip removal from the bladder or prostate.

Source facts

Action
On 07/09/21 Voluntary Medical Device Recall notices were mailed to customers. Customers were asked to take the following actions if tip detachment occurs: If the scope tube tip detaches from the telescoping tube before insertion into the patient during preparation, the Handpiece must be replaced with a new Handpiece. If the scope tube tip detaches after the Handpiece has been inserted into the patient, this is seen cystoscopically and the Handpiece must be removed and replaced with a new Handpiece. In the event that the scope tube tip separates from the Handpiece entirely, the scope tube tip may fall into the bladder or the prostate. The standard 26Fr resectoscope, used in an Aquablation procedure, will accommodate the removal of the detached scope tube tip. In addition, customers were asked to take the following actions: 1) Read the notice. 2) Stop using affected devices. 3) Remove affected lots from their regular inventory location and segregate for future return and replacement. 4) Complete and return the acknowledgement form. 5) Pass along the notice to all that need to be aware. Customers with questions were encouraged to contact Customer Service: 1.650.232.7222 or e-mail cs@procept-biorobotics.com.
Initiated
2021-07-09
Root cause
Component change control
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, IL, TX, WI, AZ, AR, MI, WV, SD, PA, CA, DC, NJ, NY, GA, IN, VA, TN and the countries of Austria, Switzerland, Germany, Spain, France, UK, Italy.
Recall number
Z-2330-2021
Recall status
Open, Classified
CDRH record ID
188425
FDA record created
2021-08-20
Recall event number
88337

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗