Skip to content

Public FDA record

Z-2331-2021: AltiVate Anatomic Shoulder System for the following Part Numbers: 1. Part # 520-08-00, Rev. C., 2. Part

Recall record Z-2331-2021 for Encore Medical, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2331-2021
Company in FDA record
Encore Medical, LP
FDA record creation date
Aug 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints have been received concerning the humeral stem failing to mate with the broach handles for a shoulder device. This could result in delay of procedure or revision surgery.

Source facts

Action
On June 28, 2021, Encore Medical (DJO) issued a "Urgent Medical Device Recall" Notification to affected consignees, via E-Mail. In addition to informing consignees about the recall devices, the firm asked consignees to take the following actions by July 19, 2021: 1. Review your stock for the affected products and lot numbers and quarantine the items until returned. 2. Communicate the affected parts, lots, and quantities that you have in stock to your sales Agent and arrange to return all affected, unused product to your sales Agent. 3. Report any adverse events related to this issue to DJO Customer Service (1-800-456-8696). Please also note that Adverse reactions or quality problems experienced with the use of this product MAY be reported to the FDAs MedWatch Adverse Event reporting program either online, by regular mail or by fax. 4. Complete the Medical Device Recall acknowledge and receipt form (including all information in the yellow-highlighted cells) with all available information and return by e-mail to julie.baker@djoglobal.com and david.evans1@djoglobal.com by July 31, 2021.
Initiated
2021-06-28
Root cause
Process change control
Distribution
US: AL,AR,AZ,CA,CO,FL,GA,HI,IL,IN,KY,LA,MD,ME,MI,MN,MO,MS,MT,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,TN,TX,UT,VA,WA,WV OUS: N/A
Recall number
Z-2331-2021
Recall status
Open, Classified
CDRH record ID
188158
FDA record created
2021-08-23
Recall event number
88223

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗