Public FDA record

Z-2331-2026: Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTI

Recall record Z-2331-2026 for HF Acquisition Co LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2331-2026
Company in FDA record
HF Acquisition Co LLC
FDA classification date
Jun 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to incorrect incubation process being performed on product.

Source facts

Action
On 04/13/2026, the firm emailed an "URGENT: Medical Device Field Correction" Letter to customers informing them that the firm has identified a laboratory testing service error associated with 1023780 SporeTraq" Mail-In Spore Testing Service, 52 Tests affecting customers using Dry Heat, Chemiclave and Ethylene Oxide autoclave types. Specifically, the incubation criteria applied to the test samples may not have complied with the above requirement for the respective autoclave type. Failure to adhere to appropriate incubation criteria impacts the validity of sterilization process followed at customer locations. Customers are instructed to: 1. Log into your OnTraq" account at ontraq.healthfirst.com. 2. A banner will appear stating ACTION REQUIRED: Review to confirm or update your autoclave types. Go to the Confirmations Page. prompting you to verify the autoclave type. 3. Verify that the autoclave type is accurate by clicking the "Go to the Confirmations Page" hyperlink or navigating to "Confirm Autoclave Type" tab - If all information is accurate as it is, click on the CONFIRM button. - If not, edit by clicking the pencil icon next to the relevant data field, enter the updated information and click on the SAVE button. - If the autoclave is no longer in use, click the DEACTIVATE button and confirm the action in the pop-up window. 4. If the above actions were completed prior to receipt of this notification via email, mail or telephone, no further action is required. For questions - contact Customer Service at 855-476-1342 or email us at customerservice@healthfirst.com.
Initiated
2026-04-13
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, and WI.
Recall number
Z-2331-2026
Recall status
Open, Classified
CDRH record ID
219495
FDA classified
2026-06-05
Recall event number
98695

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗