Z-2332-2024: MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Recall record Z-2332-2024 for Barco N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2332-2024
Company in FDA record
Barco N.V.
FDA classification date
Jul 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

Source facts

Action
Barco NV notified its sole consignee on 04/29/2024 via email. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by the customer: 1. Identify the location and amount of all Molex modules that are listed in pdf document Molex_SN_Partner . 2. Fill out and send back Response Form 1 to your Barco Key Account Manager, including the number of distributed units per country. 3. Disable or discard all used Molex modules. 4. Fill out and send back Response Form 2 to your Barco Key Account Manager after completion of the actions. For question, contact your Barco Key Account Manager.
Initiated
2024-04-24
Root cause
Other
Distribution
US: FL
Recall number
Z-2332-2024
Recall status
Completed
CDRH record ID
208118
FDA classified
2024-07-11
Recall event number
94731

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗