Z-2332-2025: Novum IQ LVP (Large Volume Pump), Product Code 40700BAXUS, infusion pump

Recall record Z-2332-2025 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2332-2025
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Aug 29, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).

Source facts

Action
Baxter issued an "URGENT MEDICAL DEVICE CORRECTION" to its consignees on 08/04/2025 via USPS first class mail. The notice described the problem, hazard involved, and actions to be Taken by customers: 1. Users should rely on their clinical judgment. When providing patient care involving high risk medications and/or critical illness, users should consider utilizing an alternate pump, if available and appropriate for the patient's treatment. 2. If an alternate pump is not available or appropriate for patient care you may continue to use the Novum IQ LVP and Novum IQ Syringe Pump consistent with the following actions: - If a blank Run screen is displayed during an infusion on the Novum IQ LVP, the user should open the door, press the tubing into load point 3, then load point 4, to turn the light green, and close the door. If a blank Run screen is displayed during an infusion on the Novum IQ Syringe Pump, the user should unload, then reload the syringe. If these actions do not resolve the blank screen, both LVP and Syringe Pump users should power-cycle the pump, select "Yes" to the new patient prompt, and reprogram the infusion. - To prevent a false motor movement system error, do not program the Novum IQ syringe pump to infuse below the Minimum Recommended Flow Rate for a syringe size of 10 mL or greater. 3. Acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. 4. Please forward a copy of this communication to any departments within your institution who use the affected product. 5. Print and post Attachment A in areas frequented by potential users. For general questions regarding this communication, or if you experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867 (select option 2, then option 2 again) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.
Initiated
2025-08-04
Root cause
Software design (manufacturing process)
Distribution
US distribution including Puerto Rico and OUS (International) Canada
Recall number
Z-2332-2025
Recall status
Open, Classified
CDRH record ID
215108
FDA classified
2025-08-29
Recall event number
97296

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗