Open original FDA record ↗Recall record
- FDA identifier
- Z-2333-2008
- Company in FDA record
- Oscor, Inc.
- FDA classification date
- Sep 24, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The O-rings are over tolerance making it hard to connect the leads to the pacemaker.
Source facts
- Action
- On January 24, 08, Oscor sent out a letter to doctors/ customers with the subject line of "Re: Market Withdrawal of Permanent Pacing Leads due to the lead connector insertion difficulty," stating they experienced some difficulties with the insertion of pacing lead connectors into the pacemaker headers. They asked that customers review the inventory list and remove all products, whose serial numbers match the serial numbers on the list and have not been implanted. Also, customers should call Oscor customer service for a Return Goods Product number and return the product back to the firm for replacement.
- Initiated
- 2008-01-29
- Root cause
- Nonconforming Material/Component
- Terminated
- 2008-11-14
- Distribution
- Nationwide Distribution
- Recall number
- Z-2333-2008
- Recall status
- Terminated
- CDRH record ID
- 71853
- FDA classified
- 2008-09-24
- Recall event number
- 48678