There is a possibility that the screw channel that holds the rod in place may be out of tolerance. When the depth of the channel is not deep enough, the rod may not be able to provide ample compression on the insert preventing the screw head from locking down to the bone screw.
Source facts
Action
Aesculap sent an "IMPORTANT PRODUCT REMOVAL NOTIFICATION" letter dated July 25, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Distribution Inventory Sheet was attached for customers to complete and return with the affected product. Contact the firm at (610) 984-9275 for questions regarding this recall.
Initiated
2012-07-25
Root cause
Nonconforming Material/Component
Terminated
2014-03-18
Distribution
Nationwide Distribution-including the states of IL, MN, NV, OK, and TX.
Recall number
Z-2333-2012
Recall status
Terminated
CDRH record ID
111846
FDA classified
2012-09-07
Recall event number
62842