- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During battery insertion there is potential for patient waveforms to move to an open receiver module or to an occupied receiver causing the intermittent inappropriate monitoring of both patients for several minutes.
Source facts
- Action
- On 07/17/08 the firm issued a "URGENT - MEDICAL DEVICE CORRECTION" letter to their customers who received the Spacelabs Medical 1400 MHz Telemetry System - Model 91341-09. The letter states the circumstances in which the device malfunctions, instructions for staff to confirm at the time of the battery insertion that the data is appearing and to remove it from service if information is not appearing. Letters will be sent with a return receipt acknowledging each customer receives the Medical Device Correction letter. Spacelabs Technical Support will respond to all technical questions and personally upgrade all switches with resisters. Contact Spacelabs Medical at 1-800-522-7025 for assistance.
- Initiated
- 2008-07-17
- Root cause
- Device Design
- Terminated
- 2011-02-16
- Distribution
- Nationwide
- Recall number
- Z-2336-2008
- Recall status
- Terminated
- CDRH record ID
- 72436
- FDA classified
- 2008-09-25
- Recall event number
- 48973