Open original FDA record ↗Recall record
- FDA identifier
- Z-2336-2025
- Company in FDA record
- Immunotech A.S.
- FDA classification date
- Aug 18, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability or rarely give falsely low patient results which might cause Permanent injury of Remote probability. Only in some (very rare) cases may the issue occur.
Source facts
- Action
- On 06/12/2025, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter determined that the Estrone RIA kit might be affected by an isolated coated tubes inhomogeneity issue causing falsely elevated results or falsely decreased results. Customers are instructed to: Beckman Coulter recommends sharing the content of the Recall Letter with their laboratory and/or medical director regarding the need to review previous patient test results. Note- All affected product have expired (expiry dates 12/31/2024 thru 04/22/2025) For questions, contact Customer Support Center: Website: http://www.beckmancoulter.com Email: imunochem@beckman.com
- Initiated
- 2025-06-12
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of NC and the countries of Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, and Netherlands.
- Recall number
- Z-2336-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214444
- FDA classified
- 2025-08-18
- Recall event number
- 97090