Z-2337-2008: HeartStarMRtX Defibrillator/Monitor Models: M3535A/M3536A HeartStart MRx with a native 256Mbyte internal

Recall record Z-2337-2008 for Philips Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2337-2008
Company in FDA record
Philips Medical Systems
FDA classification date
Sep 25, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Defective internal memory cards may exhibit one or more of the following behaviors: "Device repeatedly restarts approximately every 20 seconds "Slow device start-up of approximately 15 seconds or more. During this time, the MRx displays either a blank screen or the Philips HeartStart MRx Start-up screen prior to displaying the selected mode of operation (i.e. Monitor, AED, Pacing or Manual Defib mode). "Internal Memory Failure INOP message when attempting to print an event from the Data Management screen. "When printing from the Data Management screen, a patients Event Summary prints with some clinical events missing.

Source facts

Action
Philips issued an Urgent Medical Device Correction notification on July 17, 2008. Customers in the US received their communication package via UPS. Letters to international customers mailed via a tracked method by local Philips representatives. Customers were informed on how to locate the serial number on their devices to determine whether or not they have affected units in their inventory. Customers are asked to follow the "Action To Be Taken By Customer/User" while they await , their correction. If users continue to use prior to correction, a back up defibrillator should be readily available. Devices that exhibit repeated restart or slow start -up should be removed from service. The correction will come in the form of an internal memory card replacement. Philips will contact customers to arrange for the correction of their devices. Contact Philips Medical at 1-978-687-1501 for assistance.
Initiated
2008-07-14
Root cause
Nonconforming Material/Component
Terminated
2017-03-14
Distribution
Nationwide Canada Australia, Brazil , China, France, Germany, Great Britain, Hong Kong, Indonesia, Italy,Latvia, New Zealand,Saudi Arabia, and South Africa
Recall number
Z-2337-2008
Recall status
Terminated
CDRH record ID
72438
FDA classified
2008-09-25
Recall event number
48975

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗