Open original FDA record ↗Recall record
- FDA identifier
- Z-2338-2010
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Sep 1, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated after Beckman Coulter identified four (4) issues with the COULTER LH 500 Series System and the Unicel DxH 800 Coulter Cellular Analysis System: Issue 1: There is a potential for misidentification to occur when the system is configured in languages other than English or Chinese. The characters #,@,[,\,],`,{,|,}, or ~ are substituted or omitted in other languages. The issu
Source facts
- Action
- The customer recall notification was initiated when Product Corrrective Actions letters with attached Customer Response form were sent on the week of February 15, 2010 to all customers who purchased the COULTER LH 500 Series System and the Unicel DxH 800 Coulter Cellular Analysis System. The letters will provided the customers with an explanation of the problem identified and a work around. Customers with questions regarding the Product Corrective Action were instructed to call 800-526-7694 in the United States or Canada, or contact your local Beckman Coulter Representative.
- Initiated
- 2010-02-12
- Root cause
- Software design
- Terminated
- 2010-09-10
- Distribution
- Throughout the United States and Canada
- Recall number
- Z-2338-2010
- Recall status
- Terminated
- CDRH record ID
- 91923
- FDA classified
- 2010-09-01
- Recall event number
- 55699