Skip to content

Public FDA record

Z-2338-2021: ECHELON Oval MRI System

Recall record Z-2338-2021 for Hitachi Healthcare Americas Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2338-2021
Company in FDA record
Hitachi Healthcare Americas Corporation
FDA record creation date
Aug 24, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There was no label on the pad packaging or notification in the user manual that the table pads contained dry natural rubber which is required by the Code of Federal Regulations section 21CFR801.437.

Source facts

Action
On July 29, 2021, the firm distributed Urgent Medical Device Safety Alert letters to affected customers. Customers were informed that the anti-skid area on the bottom of the table pads contain dry natural rubber. If a patient or user inadvertently comes into contact with the anti-skid material, it may cause allergic symptoms. If this occurs, immediately seek medical care for the user or patient. The firm does not require any additional action by the user at this time. The firm will begin including revised labels and user manuals with newly manufactured systems which contain the required notification of the presence of dry natural rubber in the anti-skid area of the pads. The firm will not be taking any additional action for already distributed product. For questions, comments, or concerns, please contact: Aaron Pierce Director, QA/RA 1959 Summit Commerce Park Twinsburg, OH 44087 Phone (330) 425-1313 x2934, Fax (330) 963-0749 Email aaron.pierce@fujifilm.com
Initiated
2021-07-29
Root cause
Vendor change control
Terminated
2023-09-08
Distribution
Worldwide distribution - US Nationwide and the countries of Mexico and Brazil.
Recall number
Z-2338-2021
Recall status
Terminated
CDRH record ID
188705
FDA record created
2021-08-24
Recall event number
88452

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗