There was no label on the pad packaging or notification in the user manual that
the table pads contained dry natural rubber which is required by the Code of
Federal Regulations section 21CFR801.437.
Source facts
Action
On July 29, 2021, the firm distributed Urgent Medical Device Safety Alert letters to affected customers. Customers were informed that the anti-skid area on the bottom of the table pads contain dry natural rubber.
If a patient or user inadvertently comes into contact with the anti-skid material, it may cause allergic symptoms. If this occurs, immediately seek medical care for the user or patient. The firm does not require any additional action by the user at this time.
The firm will begin including revised labels and user manuals with newly manufactured systems which contain the required notification of the presence of dry natural rubber in the anti-skid area of the pads. The firm will not be taking any additional action for already distributed product.
For questions, comments, or concerns, please contact:
Aaron Pierce
Director, QA/RA
1959 Summit Commerce Park
Twinsburg, OH 44087
Phone (330) 425-1313 x2934, Fax (330) 963-0749
Email aaron.pierce@fujifilm.com
Initiated
2021-07-29
Root cause
Vendor change control
Terminated
2023-09-08
Distribution
Worldwide distribution - US Nationwide and the countries of Mexico and Brazil.
Recall number
Z-2339-2021
Recall status
Terminated
CDRH record ID
188706
FDA record created
2021-08-24
Recall event number
88452
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.