Information released with the product Agility did not originally contain information on compatibility with the add-on Dynamic Multileaf Collimator Apex.
Source facts
Action
ELEKTA sent an Important Field Safety Notice, FCO Ref: 20001405000 dated July 7, 2012, to current Agility users and a separate notice FCO Ref: 20001407026 for current Apex users who may be in the process of considering an upgrade to Agility. Both notices made it clear that Apex has not been integrated or classified as compatible for use with Agility. Users are advised not to use Apex with Agility.
For questions regarding this recall call 770-300-9725.
Initiated
2012-07-27
Root cause
Error in labeling
Terminated
2014-10-06
Distribution
Nationwide Distribution including DC & MD
Recall number
Z-2340-2012
Recall status
Terminated
CDRH record ID
112031
FDA classified
2012-09-10
Recall event number
62915
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.