Skip to content

Public FDA record

Z-2340-2021: Gore Molding & Occlusion Balloon Catheter, REF Catalogue Number MOB37, SN XXXXXXXX, 10 mm - 37 m 10 Fr, G

Recall record Z-2340-2021 for W. L. Gore & Associates Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2340-2021
Company in FDA record
W. L. Gore & Associates Inc.
FDA record creation date
Aug 25, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints received concerning Balloon Catheter leakage from the guidewire lumen and the y-hub during preparation resulting in either insufficient inflation or an inability to inflate.

Source facts

Action
On July 23, 2021 issue a "Urgent Medical Device Recall" notification to affected consignees via UPS. In addition to informing consignees about the recall the customer notification ask consignees to take the following actions: 1. Identify and return any unused devices within the scope of this voluntary recall. 2. Please complete and sign the enclosed CUSTOMER RESPONSE FORM and return to WLGore4731@stericycle.com within 2 weeks of receipt of this notification. 3. Please share this letter with others in your hospital or clinic as appropriate. 4. If a listed device has been used, there is no patient-follow up needed and there are no further actions needed other than informing Gore the device was used. Please indicate the used devices on the CUSTOMER RESPONSE FORM and return to WLGore4731@stericycle.com within 2 weeks of receipt of this letter. No further action is needed.
Initiated
2021-07-12
Root cause
Process change control
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AUS, CAN, CHE, DEU, DNK, ESP, FIN, FRA, GBR, GRC, HRV, HUN, IRL, ISR, ITA, JPN, MEX, NLD, NOR, POL, PRT, SMR, SWE.
Recall number
Z-2340-2021
Recall status
Open, Classified
CDRH record ID
188406
FDA record created
2021-08-25
Recall event number
88331

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.