Open original FDA record ↗Recall record
- FDA identifier
- Z-2342-2025
- Company in FDA record
- GE Medical Systems, LLC
- FDA classification date
- Aug 22, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
Source facts
- Action
- GE HealthCare will send a customer letter providing safety instructions as an immediate mitigation, to communicate the risk of additional radiation and that the equipment should only be used by personnel properly trained and authorized in the use of ionizing radiation. A user should observe proper safety practices during equipment operation as required by the operator manual. GE HealthCare will correct all affected OEC 9800 and OEC 9900 Elite systems (28 units) by visually inspecting for the presence of lead shield tape on the collimator cover and replacing the collimator cover if it is found non-compliant, at no cost to the customer.
- Initiated
- 2025-07-02
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2342-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215634
- FDA classified
- 2025-08-22
- Recall event number
- 97454