Skip to content

Public FDA record

Z-2343-2008: Ultraview SL Command Module, Model 91496 The product is intended for use with the Patient Care Monitoring

Recall record Z-2343-2008 for Spacelabs Healthcare, Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2343-2008
Company in FDA record
Spacelabs Healthcare, Incorporated
FDA classification date
Sep 19, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential for module to fail to obtain a NIBP reading on hypertension patients.

Source facts

Action
On July 22, 2008 an "Urgent MEDICAL DEVICE CORRECTION" letter was mailed to all the firm's customers. The letter states the problem at hand, when the problem would likely to occur and instructions of a way to mitigate the risk, until the modules can be recalibrated . The firm's Customer Connectivity Center will confirm all affected customers have been notified. Spacelabs Technical Support will be recalibrating all modules personally and ensuring all corrections have been applied to all affected customers of record. Contact Spacelabs Healthcare at 1-800-522-7025 for assistance.
Initiated
2008-07-22
Root cause
Process control
Terminated
2011-07-25
Distribution
The product is distributed nationwide and internationally to medical facilities.
Recall number
Z-2343-2008
Recall status
Terminated
CDRH record ID
72427
FDA classified
2008-09-19
Recall event number
48759

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗