Z-2343-2018: 26-Lead Transfer Set for use with APEX Compounding System, Product Catalog Number 2112550

Recall record Z-2343-2018 for B. Braun Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2343-2018
Company in FDA record
B. Braun Medical, Inc.
FDA record creation date
Jul 3, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for valves on the 26-Lead Transfer Set to be incorrectly oriented. When a transfer set with an incorrectly oriented valve is installed on the system, there is potential for unintended admixture.

Source facts

Action
Urgent Medical Device Correction - Recall Notification letters were distributed on 6/4/18. The letters instruct customers to perform the following: 1. Review this Medical Device Correction Notification in its entirety and ensure that all users of the above mentioned products in your organization and other concerned persons are informed about this medical device correction - recall notification. If you are a distributor, please forward this recall notification to your customers. 2. Perform a visual inspection of valve orientation prior to installation onto APEX Compounding System in accordance with the Visual Inspection Instructions contained in Attachment 1. 3. If any misaligned valves are identified, do not use the transfer set. Please contact Customer Support at 1-800-227-2862 for product return and replacement. 4. Utilizing the attached "Medical Device Correction - Recall Notification Acknowledgement" form, please acknowledge that you have received and reviewed this information. 5. Return the completed "Medical Device Correction - Recall Notification Acknowledgement" form to B. Braun Medical Inc Quality Assurance department by faxing or e-mailing the form within two (2) weeks of receipt of this notice. It is important this form is returned to B. Braun Medical Inc can meet regulatory requirements of the United States Food and Drug Administration.
Initiated
2018-06-04
Root cause
Under Investigation by firm
Terminated
2020-01-13
Distribution
The products were distributed to the following US states: AL, CA, CO, CT, FL, GA, IL, LA, MA, MD, MI, MN, MO, NC, NJ, NV, OH, PA, TX, and WA.
Recall number
Z-2343-2018
Recall status
Terminated
CDRH record ID
164915
FDA record created
2018-07-03
Recall event number
80231

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗