Z-2345-2025: Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable

Recall record Z-2345-2025 for Dynex Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2345-2025
Company in FDA record
Dynex Technologies, Inc.
FDA classification date
Aug 22, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.

Source facts

Action
On July 16, 2025, URGENT FIELD SAFETY NOTICE letters were emailed to customers. Action to Be Taken by the User: - Upon receipt, Quarantine affected devices and return or destroy them. - Customers have been provided a Field Safety Notice Customer Reply Form. Please complete this form, sign it and return to Dynex as soon as possible to acknowledge that you have received this information and have destroyed or returned to Dynex the impacted 67910 Agility Sample Tips Lot 863366. Transmission of this Field Safety Notice This notice needs to be passed on all those who need to be aware within your organisation or to any organisation where the potentially affected devices have been transferred. (As appropriate) Please transfer this notice to other organisations on which this action has an impact. (As appropriate) Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-related incidents to the manufacturer, distributor or local representative, and the national Competent Authority if appropriate, as this provides important feedback..*
Initiated
2025-07-16
Root cause
Labeling Change Control
Distribution
Domestic: CA, FL, NC, NJ, NY;
Recall number
Z-2345-2025
Recall status
Open, Classified
CDRH record ID
215221
FDA classified
2025-08-22
Recall event number
97319

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗