Open original FDA record ↗Recall record
- FDA identifier
- Z-2346-2024
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA classification date
- Jul 11, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Source facts
- Action
- On June 3, 2024 IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to consignees. Fresenius Medical Care is requesting replacement of affected Blood Pump Rotors in 2008K@HOME systems due to the patients possibly being unaware of this component failing. To order for a replacement Blood Pump Rotor (Part Number, F40015481), contact Fresenius Medical Care Customer Service at 800-227-2572, OPTION 3, o Please have Blood Pump Rotor Part Number and Lot Number available when requesting a replacement. CUSTOMER: 1. Please provide this letter to your Biomedical Technician. 2. Report any complaints or adverse events to product.complaints@fmc-na.com or FDA MedWatch at https://www.fda.gov/safety/medical-product-safety-information/medwatch-forms-fda-safety-reporting. 3. Provide this Customer Notice to all those who need to be aware within your organization. 4. Complete and return the attached reply form. For any clinical or medical questions, please contact the Global Medical Information and Education Office at 1-855-616-2309.
- Initiated
- 2024-06-03
- Root cause
- Component design/selection
- Distribution
- Domestic: Nationwide Distribution. International: Canada.
- Recall number
- Z-2346-2024
- Recall status
- Open, Classified
- CDRH record ID
- 208290
- FDA classified
- 2024-07-11
- Recall event number
- 94775