Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo
Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8
monitors were manufactured with an underrated circuit breaker due to an error where an associate
pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory
locations hav
Source facts
Action
TERUMO Cardiovascular Systems communicated with their affiliate, Terumo Dubai, in 2009 at the time the complaints were reported. No further communication was warranted.
For questions regarding this recall call 734-741-6173.
Initiated
2009-08-12
Root cause
Mixed-up of materials/components
Terminated
2013-01-29
Distribution
Worldwide distribution: International only-country of: Terumo Dubai, UAE
Recall number
Z-2348-2012
Recall status
Terminated
CDRH record ID
110221
FDA classified
2012-09-11
Recall event number
62321
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.