Z-2348-2019: Thoratec HeartMate 3 LVAS Implant Kit, Sterile, REF 106524/106524US, RX only, Left Ventricular Assist Sys

Recall record Z-2348-2019 for Thoratec Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2348-2019
Company in FDA record
Thoratec Corp.
FDA record creation date
Aug 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.

Source facts

Action
On March 30, 2019 all consignees were hand delivered the "Urgent Medical Device Recall" letter. The recall letter stated the following: 1. Abbott is recalling 208 HeartMate 3 Outflow Graft units globally (87 within the US). An Abbott Representative will be contacting you within the coming weeks to identify the specific units that are impacted in your inventory and replace them. 2. In the interim, continued implants with the current HeartMate 3 Outflow Grafts can still be conducted with standard implant procedures by following the HeartMate 3 Instructions for Use. However, in addition to following the existing instruction of checking for blood leaks and bleeding at the Outflow Graft When the flow through the blood pump is satisfactory, we now recommend that adequate performance of the Screw Ring attachment to the pump be evaluated before the device is brought to the surgical field. If the Screw Ring does not function properly, the Outflow Graft should be replaced with a backup Outflow Graft. Patient Management Recommendations 3. For patients implanted with the HeartMate 3 Left Ventricular Assist System (LVAS), Abbott confirms there is no risk due to this issue. Patients should be managed per standard clinical practice. 4. Please complete the acknowledgement form included in this packet and return to Abbott. If you are a consignee of this letter within your organization, please notify all users of the device within your organization. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report Online: www.fda.gov/medwatch/report.htm " Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed from, or submit by fax to 1-800-FDA-0178. 6. If you have questions,
Initiated
2019-03-30
Root cause
Process design
Terminated
2020-08-24
Distribution
Worldwide distribution. US nationwide, Australia, Austria, Brazil, Canada, Cayman Islands, Chile, Colombia, Egypt, France, Germany, Greece, Singapore, Iran, Italy, Kuwait, Lebanon, Romania, Switzerland, Thailand, United Kingdom.
Recall number
Z-2348-2019
Recall status
Terminated
CDRH record ID
174929
FDA record created
2019-08-22
Recall event number
83357

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗