Open original FDA record ↗Recall record
- FDA identifier
- Z-2349-2010
- Company in FDA record
- AMD Lasers, LLC
- FDA classification date
- Sep 10, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Product is noncompliant with the Federal laser product performance standard in certain requirements. The ON/OFF key on some of the Picasso/Picasso Lite lasers could be removed with force while the key was in the ON position.
Source facts
- Action
- Letters, dated April 30, 2010, addressed to Dental Professionals were sent identifying the affected product and reason for the recall. The letter asked customers to check their units and, if applicable, AMD Lasers would replace the key assembly on their lasers. The service would be provided free of charge including round trip shipping charges to the corporate office. Customers are to call to schedule the service and obtain a RMA number and shipping instructions. Questions should be directed to 1-866-999-2635 or support@amdlasers.com.
- Initiated
- 2010-04-15
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2011-04-21
- Distribution
- Worldwide Distribution
- Recall number
- Z-2349-2010
- Recall status
- Terminated
- CDRH record ID
- 93241
- FDA classified
- 2010-09-10
- Recall event number
- 56330