Skip to content

Public FDA record

Z-2349-2021: Vascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler (USV2). Intended Use: Vasc

Recall record Z-2349-2021 for Oakworks Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2349-2021
Company in FDA record
Oakworks Inc
FDA record creation date
Aug 25, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Weld failure- the bracket attaches the tabletop tilting actuator to the tabletop frame. If the bracket weld fails, the tabletop tilting action is non-functional

Source facts

Action
Oakworks issued Urgent Medical Device Recall letter dated 6/29/21. Letter states reason for recall , health risk and action to take: Actions to be taken are to immediately examine your inventory of OAKWORKS Inc., Ultrasound Vascular Table without Fowler (V1) and Ultrasound Vascular Table with Fowler (V2) to identify and locate all affected product(s). Do not distribute these affected product(s). Please quarantine the table(s). For tables that have been in use, they remain fully functional. Implement the field action as described below for the product listed in the Product and Distribution Information. Enclosed is a Medical Device Advisory Notice Response Form for each product we believe you have. Please complete this form in its entirety and return to OAKWORKS Inc., within 48 hours of receipt of this advisory notice via any of the options listed below. If you have further distributed this product, immediately contact your customer(s) with a copy of this letter and all applicable response forms. Then, provide OAKWORKS Inc. with the contact information for your customer(s) to whom this product(s) has been distributed by using the Medical Device Advisory Notice Response Form within 48 hours. Please note that you are required by the United States Food and Drug Administration to contact your customer(s) who are currently in possession of the product and inform them of this Medical Device Advisory Notice. MAIL: Oakworks Inc. EMAIL: qualitycare@oakworks.com Attn: Quality Care MDAN Response FAX: (717) 235-6798 923 East Wellspring Rd. New Freedom, PA 17349 What OAKWORKS Inc. will do next: Upon receipt of the Medical Device Advisory Notice Response Form(s), and to minimize the downtime of your OAKWORKS Ultrasound Vascular Table (V1) or (V2), we will setup a service visit to replace the Tilt Actuator Bracket on the table(s). The correction installation takes approximately 30 minutes. The replaced bracket will be disposed of by the service provider. If you have any quest
Initiated
2021-06-29
Root cause
Process control
Terminated
2023-10-05
Distribution
Nationwide
Recall number
Z-2349-2021
Recall status
Terminated
CDRH record ID
188513
FDA record created
2021-08-25
Recall event number
88365

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗