Open original FDA record ↗Recall record
- FDA identifier
- Z-2351-2010
- Company in FDA record
- Abbott Laboratories
- FDA classification date
- Sep 2, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks.
Source facts
- Action
- Product Recall letters were sent on 04/16/2010. The letters identified the affected product, explained the reason for recall, the patient impact/user safety, stated the necessary actions, and contact information. Users are to discontinue use until the Waste Chamber(s) have been replaced and follow the instructions in the Hazards section. In addition, appropriate protective equipment needs to be worn along with following their recommended laboratory procedures dealing with spills. Users are to contact their local Abbott Customer Support rep. US customers should be direct their questions to Customer Support.
- Initiated
- 2010-04-16
- Root cause
- Nonconforming Material/Component
- Terminated
- 2010-12-06
- Distribution
- Worldwide Distribution -- CA, CT, ID, IN, and MA; and, countries of Australia, Brazil, Canada, China, Colombia, Mexico, Puerto Rico, Singapore, and Uruguay.
- Recall number
- Z-2351-2010
- Recall status
- Terminated
- CDRH record ID
- 93216
- FDA classified
- 2010-09-02
- Recall event number
- 56309