Reason stated in the source
potential contamination with potassium hydroxide may have an impact when performing immunohistochemistry testing. It is unknown whether the contamination will affect tissue antigenicity. If it were to affect tissue antigenicity, the outcome would be risk of false positive or false negative test results which may lead to serious injury. There is the possibility the patient may need to be re-biopsied.
Source facts
- Action
- On 6/13/2016, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected customers via certified mail with instructions for identifying and disposing of the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.
- Initiated
- 2016-06-13
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-11-08
- Distribution
- Distributed in the states of FL, PA, NY, MD, CA, and AZ.
- Recall number
- Z-2351-2016
- Recall status
- Terminated
- CDRH record ID
- 147901
- FDA record created
- 2016-08-05
- Recall event number